These water-cooled high-temperature extensometers are designed for measuring axial strain in high temperature tension, compression, and through-zero fatigue testing applications up to 1200°C (2200°F). Water cooling is more effective at keeping the extensometer body at a constant temperature and reducing noise in the data measured.
Test types
Test specimens
Test standards
| Metric Units | |||||||
|---|---|---|---|---|---|---|---|
| MODEL NUMBER | OPTIONS | GAGE LENGTH | MEASURING RANGE (% STRAIN) | MAXIMUM TEMPERATURE | CONTACT FORCE | CERAMIC EXTENSION RODS INCLUDED | PART NUMBER |
| 632.51F-04 | 006,080 | 12 mm | +20% / -10% | 1200°C | 780 g | Qty 2, V-chisel Edge Rods, 3.50 mm (.138 in) rod diameter, 85 mm (3.35 in) long | 047-402-651 1 |
| 632.51F-01 | 006,077 | 25 mm | +15% / -10% | 1200°C | 400 g | Qty 2, V-chisel Edge Rods, 3.50 mm (.138 in) rod diameter 94 mm (3.7 in) long | 047-402-652 1 |
| US Customary Units | |||||||
| MODEL NUMBER | OPTIONS | GAGE LENGTH | MEASURING RANGE (% STRAIN) | MAXIMUM TEMPERATURE | CONTACT FORCE | CERAMIC EXTENSION RODS INCLUDED | PART NUMBER |
| 632.51E-04 | 006,080 | 0.5 in | +20% / -10% | 2200°F | 780 g | Qty 2, V-chisel Edge Rods, 3.50 mm (.138 in) rod diameter, 85 mm (3.35 in) long | 047-402-650 1 |
| 632-51F-01 | 006,077 | 1.00 in | +15% / -10% | 1200°F | 400 g | Qty 2, V-chisel Edge Rods, 3.50 mm (.138 in) rod diameter, 94 mm (3.7 in) long | 047-402-653 1 |
Specifications are taken from the official MTS product page and may change; please confirm with our team before ordering.
MTS Systems
From complex thermomechanical fatigue studies to standard tension and compression tests, MTS Landmark systems deliver accurate, repeatable results.
MTS Systems
These air-cooled high-temperature extensometers are designed for measuring axial strain in high temperature tension, compression, and through-zero fatigue testing applications up to 1200°C…
MTS Systems
Series 647 All-Temperature Grips can be operated within an environmental chamber at temperatures up to 1000°F (540°C).
MTS Systems
MTS Model 680 High-Temperature Grips provide accurate results when testing temperature-sensitive materials and minimize risk from thermal stress on the specimen.